Refurbished Agilent 1100 Series HPLC SystemThe Agilent 1100 Series HPLC System remains one of the most widely deployed high-performance liquid chromatography platforms across pharmaceutical, chemical, and academic laboratories globally. Launched as a modular, scalable system, the Agilent 1100 set a benchmark for reliab...
Need Expert Advice?
Call us: +91 9958700332
Product Details
Refurbished Agilent 1100 Series HPLC System
The Agilent 1100 Series HPLC System remains one of the most widely deployed high-performance liquid chromatography platforms across pharmaceutical, chemical, and academic laboratories globally. Launched as a modular, scalable system, the Agilent 1100 set a benchmark for reliability and detector flexibility that continues to make it relevant in active lab environments today. At Spectrachrome, we supply fully refurbished Agilent 1100 Series HPLC systems that are performance-verified, properly configured, and ready for immediate analytical use at a price point that reflects responsible procurement, not compromise on quality.
About the Agilent 1100 Series HPLC
- The Agilent 1100 Series is a fully modular HPLC platform built around a quaternary pump capable of blending up to four solvents simultaneously, enabling both isocratic and complex gradient separations. Its autosampler delivers injection volumes from 0.1 to 100 µL with carry-over performance that meets the demands of trace-level pharmaceutical analysis. The system integrates seamlessly with the Agilent 1100 diode array detector (DAD) for full-spectrum UV-Vis data acquisition, or with variable wavelength detectors (VWD), fluorescence detectors (FLD), and refractive index detectors (RID) depending on the analytical method.
- ChemStation software provides complete system control, data acquisition, method management, and regulatory-compliant audit trail functionality a critical requirement for pharmaceutical laboratories operating under 21 CFR Part 11 or Schedule M compliance frameworks.
Key Technical Specifications-
| Component / Feature | Specification |
| Pump Type | Quaternary gradient pump (G1311A) |
| Flow Rate Range | 0.001 – 10.000 mL/min |
| Operating Pressure | Up to 400 bar (5,800 psi) |
| Autosampler Injection Range | 0.1 – 100 µL |
| Column Thermostat Range | 10°C below ambient to 80°C |
| Detector Options | DAD, VWD, FLD, RID |
| DAD Wavelength Range | 190 – 950 nm |
| Software | Agilent ChemStation |
| Communication | LAN, GPIB, RS-232 |
Spectrachrome Refurbishment Process
Every Agilent 1100 Series HPLC system supplied by Spectrachrome undergoes a structured, module-by-module reconditioning process. We restore each unit to its original operating specification — not just its appearance.
Our process includes:
• Quaternary pump inspection and seal replacement — pump heads disassembled, seals and pistons inspected, flow accuracy calibrated at multiple set points.
• Autosampler needle and injection valve service — needle, rotor seal, and injection valve inspected and replaced as required to eliminate carry-over.
• Column oven calibration — temperature uniformity verified across the full operating range.
• DAD/VWD lamp performance check — deuterium lamp output measured against specification; replaced if below threshold.
• ChemStation connectivity and method verification — full software communication confirmed with method run and data export validation.
• System suitability testing — retention time reproducibility, peak symmetry, and baseline noise tested using reference standards before dispatch.
Every unit ships with a performance qualification report. You are not buying potential you are buying a verified result.
Ideal Applications for the Agilent 1100 Series HPLC
| Pharmaceutical QC | Method validation, assay of APIs and finished dosage forms, impurity profiling, and dissolution testing under ICH Q2(R1) and USP compliance requirements. |
| Bioanalytical Research | Plasma protein quantification, metabolite profiling, and pharmacokinetic studies requiring high-sensitivity gradient separations with fluorescence or DAD detection. |
| Academic Research | Postgraduate and doctoral research environments needing a dependable gradient HPLC platform for natural product isolation, synthesis monitoring, and method development. |
| Chemical & Industrial QC | Purity testing of chemical intermediates, raw material qualification, and process analytical monitoring across batch production. |
| Environmental Testing | Pesticide residue analysis, PAH determination, and water quality monitoring using validated regulatory methods. |
Why the Agilent 1100 Series Remains a Smart Buy
- The Agilent 1100 HPLC has one of the largest installed bases of any HPLC platform in the world. This means spare parts are readily available, service engineers are familiar with the system, and validated methods written for the 1100 Series can be transferred without revalidation in most cases. For laboratories running established pharmacopoeial or in-house methods, this continuity has direct regulatory and operational value.
- A refurbished Agilent 1100 Series HPLC from Spectrachrome delivers full gradient capability, multi-detector flexibility, and ChemStation software integration at 40–60% below the cost of a comparable new system without sacrificing the analytical performance your laboratory depends on.
PAN India Delivery & Post-Sale Support
Spectrachrome delivers across India including Delhi, Mumbai, Bengaluru, Hyderabad, Pune, Chennai, Kolkata, and Ahmedabad. On-site installation, ChemStation setup, and initial method run support are available on request.
We also provide:
• Spare parts supply — pump seals, autosampler needles, rotor seals, DAD lamps, and detector flow cells.
• Annual Maintenance Contracts (AMC) covering scheduled servicing and calibration.
• ChemStation software installation, qualification, and user training.